CsI and Si pedestal
Si pedestals procedure
----------------------
- On the trigger / selecteur interface (indra_trigger):
- In "Modes" box,
put "physique" button in red
put "generateur" button in green
- Click on "Declenchement generateur",
Put "generateur test" button in green
and all other buttons in red
- Click on "Valider Ecriture VXI"
- Under GAG cmd > type : VME EXECUTE START_PIED_SI
- Start the acquisition and check the MRQ cable
To go back to normal mode:
- On the trigger / selecteur interface:
- In "Modes" box click on the modes you wish and check
the "Declenchement generateur"
Under GAG cmd > type : VME EXECUTE START_PIED_SI_RESTORE
CsI pedestals procedure
-----------------------
- On the trigger / selecteur interface (indra_trigger):
- In "Modes" box,
put "physique" button in red
put "generateur" button in green
- Click on "Declenchement generateur",
Put "generateur test" button in green
and all other buttons in red
- Change "Fast Trigger Test Retard" from 730 ns to 200 ns.
- Click on "Valider Ecriture VXI"
- Under GAG cmd > type : VME EXECUTE START_PIED_CSI
- Start the acquisition and check the MRQ cable
To go back to normal mode:
- On the trigger / selecteur interface:
- In "Modes" box click on the modes you wish and check
the "Declenchement generateur"
- Change "Fast Trigger Test Retard" from 200 ns to 730 ns.
- Under GAG cmd > type : VME EXECUTE START_PIED_CSI_RESTORE