PEDESTRIAL PROCEDURE
Septembre 1998
Si (Chio are also done) pedestals procedure
1- On the trigger interface ("SELECTEUR" graphical window):
- In "Modes" box, put "physique" button in red and "generateur" button in green
- Click on "Declenchement generateur", put "generateur test" button in green and all other buttons in red
- Click on "Valider Ecriture VXI"
2- Start the 1D spectra :
- Under GAG cmd > Type : VME EXECUTE START_PIED_SI
3- On the inspection panel:
- Set the MRQ to level 1
4- Start the acquisition
- Under GAG cmd > Type : acq start
To go back to normal mode:
1 - On the trigger / selecteur interface:
- In "Modes" box click on the modes you wish and check the "Declenchement generateur"
2 - Under GAG cmd > type :
- Type VME EXECUTE START_PIED_SI_RESTORE
3- On the inspection panel:
- Set the MRQ back to level 0.
CsI pedestals procedure
1- On the trigger / selecteur interface (indra_trigger):
- In "Modes" box, put "physique" button in red and "generateur" button in green
- Click on "Declenchement generateur", Put "generateur test" button in green and all other buttons in red
- Change "Fast Trigger Test Retard" from 730 ns to 200 ns and type RETURN !.
- Click on "Valider Ecriture VXI"
2- Demarrer les spectres monodim :
- Under GAG cmd > Type : VME EXECUTE START_PIED_CSI
3- On the inspection panel:
- Set the MRQ to level 1
4- Start the acquisition
- Under GAG cmd > Type : acq start
To go back to normal mode:
1- On the trigger / selecteur interface:
- In "Modes" box click on the modes you wish and check the "Declenchement generateur"
- Change "Fast Trigger Test Retard" from 200 ns to 730 ns and type RETURN !.
2 - Under GAG cmd >
- Type : VME EXECUTE START_PIED_CSI_RESTORE
3- On the inspection panel:
- Set the MRQ back to level 0.
Olivier Lopez, David Gourio