Septembre 1998
First part : forward rings :
1) Install the alpha source on the INDRA target holder and put it in "beam position", the active face forward (looking at rings 1-13)
2) Turn off all the CsI.
3) In the "Selecteur" panel :
5) Due to the difference of solid angle size between the rings, ring 8-9 will trigger the acquisition more often than for example ring 1. To balance the statistic, turn of the ring 8-9 when the number of counts is high enough. Then latter on, when the ring 6-7 has also a good statistic, turn it off too. And so on until ring 1.
6) Turn on CsI 10-12 and ring 10-11 and 12-13.
Second part (backwards rings) :
2) Turn on all the CsI
3) Turn the alpha source of 180 degres to the backward angles.
4) Do in the same way than for the forward rings (after a while, turn off the rings 13, then the ring 14-15).